Chapter 2: Theory of Operation 45
a constant pressure. You can view the current sheath pressure setting using the
Cytometer > Sheath Pressure command (Figure 2-2).
Figure 2-2 Sheath pressure level
After fluidics startup, sheath flow is controlled using the Stream button in the
Breakoff window (Figure 2-3). When clicked, the button changes from a red “X”
to a green checkmark, and sheath flows through the cuvette flow cell at the rate
that is specified in the Sheath Pressure window. (See Drop Formation on page 61
for a complete description of the Breakoff window.)
Figure 2-3 Stream control in Breakoff window
As a general rule, the sheath pressure level is set by choosing a Sort Setup mode
from the Sort menu, rather than by adjusting the Sheath Pressure control. Each
Sort Setup mode is optimized at a preset sheath pressure. If you change the sheath
pressure, a multitude of other values will be affected and need updating,
including the drop drive frequency, drop charge levels, laser delay, area scaling
factor, and other values. For more information, see Sort Setup on page 101.