292 BD FACSAria II User’s Guide
4 Lift off the ULPA filter and attached tubing from the evacuator and dispose
of both the filter and the tubing.
5 Insert the new ULPA filter into the evacuator filter well (Figure C-8).
Figure C-8 Inserting new filter
6 Push down on the filter to ensure that it is seated against the bottom of the
filter chamber (Figure C-9 on page 293).
All biological specimens and materials coming into contact with them can
transmit potentially fatal disease. Handle the ULPA filter, attached tubing,
and all instrument hardware as if capable of transmitting infection. Dispose
of waste using proper precautions and in accordance with local regulations.
Wear suitable protective clothing, eyewear, and gloves.