164 BD FACSAria II User’s Guide
Setting Up for Bulk Sorting
This section describes how to set up the streams for two- or four-way sorting. For
sorting using the ACDU option, see Setting Up for Sorting Into a Plate or Slide on
page 179.
1 Install collection tubes in the required collection device.
Collection tube holders are available for 1-mL microtubes, 12 x 75-mm tubes,
and 15-mL centrifuge tubes. For compatible tubes, see Labware on page 269.
2 Install the collection tube holder onto the cytometer.
a Close the sort block door and open the sort collection chamber door, if
needed. The sort block door must be closed to install the tube holder.
b Slide the holder into the slotted fittings below the sort aspirator drawer,
then close the sort collection chamber door.
3 Turn on the deflection plates.
Click the Voltage button in the Side Stream window (Figure 5-1 on
page 165). The voltage warning light illuminates, indicating that the plates
are charged.
Any cytometer surface that comes in contact with biological specimens can
transmit potentially fatal disease. Use universal precautions when handling
sorting hardware. Wear suitable protective clothing and gloves.
Before installing the collection tube holder, ensure that an O-ring is
installed in the groove between the two sections of the tube holder. The
O-ring minimizes the chance of aerosols escaping. It can be found in the
accessory kit (Catalog No. 337897).