Chapter 6: Shutdown and Maintenance 199
4 Disconnect the bleach fluid line from the DI water port, and connect it
back to the bleach tank port, then reconnect the fluid line to the DI water
port, and click Done.
5 Disconnect the fluid line from the sheath tank (at the output side of the
sheath filter) and connect it to the fluid out port on the side of the fluidics
cart (see Figure 6-5). Click Done.
The system cleaning takes approximately 20 minutes.
Figure 6-5 Moving the sheath fluid line
6 Disconnect the fluid line from the fluid out port on the side of the fluidics
cart and connect it to a new 0.2-µm sheath filter.
7 Remove the old sheath filter and connect a new 0.2-µm sheath filter to the
liquid port of the sterilized sheath tank.
8 To complete the process, select one of the two options.
Figure 6-6 Finishing the aseptic sort prep process
On
(
I
)
Auxiliary
Air Supply
(
J
)
Off
(
O
)
Air In
(
F
)
sheath fluid line
connected to fluid
out port