236 BD FACSAria II User’s Guide
No stream when Stream
control clicked
Sheath tank low or empty Refill the sheath tank. See
Refilling the Sheath Tank on
page 123.
Note that when the empty tank
warning message is not dismissed
after 15 minutes, the stream shuts
off automatically.
Air in sheath line Prime the sheath tank. See Prime
After Tank Refill on page 196.
Air in sheath filter Purge the sheath filter. See Purging
the Sheath Filter on page 200.
Dry filter Open the bleeder valve to purge
the filter. See Purging the Fluid
Filters on page 200.
Fanning around center
stream
Nozzle inserted improperly Re-insert the nozzle. Push it gently
all the way forward without
rocking it from side to side.
Unstable stream Debris in flow cell or nozzle Remove the nozzle and run the
stream with no nozzle in place for
approximately 10 seconds. (Click
the Stream control on, then off.)
Sonicate the nozzle and re-install
it.
Fluid line connected to
ethanol shutdown tank
Move fluid and air lines to sheath
tank, then perform Fluidics
Startup.
Troubleshooting the Stream (continued)
Observation Possible Causes Recommended Solutions