144 BD FACSAria II User’s Guide
Data Collection
Before you record data for a sample, cytometer settings should be optimized to
position the cells of interest on scale for scatter and fluorescence parameters. In
the previous section, application settings were created by taking into
consideration the following:
•FSC and fluorescence area scaling
FSC and SSC voltages
•FSC threshold
•Fluorescence PMT voltages
In this section, the Application Settings will be applied and compensation will be
calculated before collecting test data.
The following sections describe how to use previously optimized application
settings for a 4-color experiment. See Application Settings on page 136 for more
information.
Compensation will be automatically calculated using the Compensation Setup
feature. For more information about this feature, refer to the BD FACSDiva
Software Reference Manual. If you are performing compensation manually, not
all steps will apply.
Setting Up the Workspace
Before you begin optimizing settings, it is important to verify the cytometer
configuration and create an experiment containing appropriate parameters for
the assay.
1 Select Cytometer > View Configurations, and verify that the current
configuration is appropriate for the type of sample you are running.
2 Verify that there is a valid performance check for the configuration.