116 BD FACSAria II User’s Guide
2 Verify that the air and fluid lines are disconnected from the ethanol tank
and connected to the sheath tank, then click Done. See Figure 4-3.
Figure 4-3 Air and fluid lines connected to sheath tank
3 Verify that a closed-loop nozzle is installed in the flow cell, then click Done.
The fluidics startup process starts and the progress is displayed at the
bottom of the dialog.
Caution: Do not run fluidics startup with the air and fluid lines connected
to the ethanol shutdown tank. This can cause damage to the system.
Always verify that the lines are connected to the sheath tank.
air line
fluid line
ethanol shutdown
tank
sheath tank
sheath filter
Vedere la pagina 115