Chapter 4: Running Samples 117
4 Remove the closed-loop nozzle from the flow cell, then click Done.
5 Insert the correct nozzle size in the flow cell.
a Verify that the O-ring is installed in the nozzle.
b Make sure the top side of the nozzle is facing up as you insert the
nozzle.
c Push the nozzle all the way back into the flow cell.
d Turn the nozzle-locking lever clockwise to the 12:00 position.
6 Click OK to complete the process.
7 When fluidics startup is complete, select Sort > Sort Setup and verify that
the setup mode matches the nozzle size.
For information about Sort Setup modes, see Sort Setup on page 101.