Aria 8.0 Guida Utente Pagina 28

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 27
28 BD FACSAria II User’s Guide
Nozzle
The BD FACSAria II cytometer is provided with four nozzle sizes (70, 85, 100,
and 130 Pm) to accommodate a variety of particle sizes. The nozzle is keyed to a
fixed position at the lower end of the cuvette. Because the nozzle is below the
interrogation point, optical alignment is not affected when the nozzle is changed.
Figure 1-11 Nozzle
When installing or removing the nozzle, make sure the O-ring is in place before
proceeding. Look at the O-ring in the nozzle with a magnifier or under a
microscope to confirm it is positioned correctly.
Refer to these sections for more information on the nozzle:
•Changing Nozzles on page214
Cleaning the Nozzle on page 216
O-ring
Vedere la pagina 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 343 344

Commenti su questo manuale

Nessun commento