128 BD FACSAria II User’s Guide
Figure 4-9 Waste container details
1 Disconnect the waste container’s sensor and fluid line connectors from their
respective ports on the fluidics cart.
2 Remove the disposable waste cap (large-sized cap) and attached trap from
the container. Place the assembly on the bench label-side up.
The waste container can become pressurized when the cytometer is
running. Always disconnect the container from the fluidics cart before you
empty it. Wait at least 1 minute for pressure to dissipate before you open
the container.
Do not wet the cap on top of the trap. If you see liquid inside the trap,
remove the drain plug and fully drain the liquid before you replace the plug
(Figure 4-10 on page 129).
waste container
trap
waste ports
(behind handle)
disposable cap
sensor
waste fluid lines